Drum-type Gasmeters TG-Series (Wet-Type)

Packing Liquids CalRiX

1. General

CalRiX is a completely synthetic fluid on a fluorine base. It is of low molecular weight, colourless, clear and odourless.

More detailed information about the chemical composition and molecular structure of this fluid can be obtained upon request.

2. Application

CalRiX can be used with all Ritter Drum-type Gas Meters (Wet-type Gas Meters). Because CalRiX is completely inert to most gases including oxygen, it can be used as a Packing Fluid when water or paraffin oil are not suitable. For example, when the gas needs to remain dry, and when the gas is highly reactive to water or paraffin oil. It is appropriate for use with such gases as:

  • Butane
  • Carbon Dioxide
  • Carbon Tetrachloride
  • Carbon Tetrafluoride
  • Chlorine
  • Deuterium
  • Fluorine
  • Helium
  • Hydrogen chloride
  • Hydrogen Fluoride
  • Methane
  • Nitrogen trichloride
  • Nitrogen trifluoride NF3
  • Oxygen
  • Phosphine
  • Propane
  • Silane
  • Sulphur Hexafluoride

3. Advantages

  • Extremely resistant even against highly aggressive gases because of CalRiX's fluorine base,
  • less evaporation than water due to its lower vapour pressure, resulting in greater stability of the Packing Fluid level and in more consistent measurement result,
  • very smooth rotation of the measuring drum in the Gas Meter because of CalRiX's high density and low surface tension.

4. Properties

Viscosity:

-20

°C

11,7

mm2/sec (= cSt)

 

20

°C

2,7

mm2/sec

 

25

°C

2,4

mm2/sec

 

100

°C

0,7

mm2/sec

Density:

20

°C

1,80

g/ml

 

100

°C

1,64

g/ml

Vapour pressure:

20

°C

0,4

mbar

 

100

°C

30,8

mbar

 

120

°C

65,6

mbar

Working Temp. Range:

-20

°C

to 190

°C

Boiling Point::

200

°C

 

 

Pour point:

-85

°C

 

 

Solubility of Water:

14

ppm

 

 

Solubility of Air:

26

cm3 gas per 100 cm3 liquid

Volatility:

34.4

% in 22 Stunden bei 66°C

Appearance:

Clear, odourless, colourless fluid

5. Solubility of Gases (Bunsen coefficient at room temperature)

Butane

8.5

Boron Trichloride

13.1

Boron Trifluoride

0.22

Carbon Dioxide

1.2

Carbon Tetrachloride

52.6

Carbon Tetrafluoride

0.68

CFC 114

14.9

CFC 12

4.25

CFC 133a

13.9

CFC 134a

4.7

CFC 21

13.1

CFC 22

4.86

Chlorine

3.19

Deuterium

0.10

Esafluoroethane

2.12

Fluorine

0.20

Helium

0.08

Hydrochloric acid

0.806

Hydrogen

0.10

Methane

0.17

Nitrogen

0.19

Nitrogen Trichloride

0.83

Nitrogen Trifluoride approx.

0.9

Oxygen

0.29

Phosphine

0.67

Propane

3.8

Silane

0.36

Sulphur Hexafluoride

3.5

The Bunsen coefficient [N ml/ml] is the volume of gas, reduced to Normal condition (1013 mbar, 0°C), dissolved in the volume unit of fluid.

Subject to alterations

 